1 MC19-1.20
Operating Instructions
Temperature Controller
Type 3508
English
Ack
3508
1200
23
RUN/HOLD
RUN/HLD
OP1
°C
%
96.0
A/MAN
Page view 0
1 2 3 4 5 6 ... 23 24

Summary of Contents

Page 1 - Type 3508

1 MC19-1.20 Operating Instructions Temperature Controller Type 3508 English Ack

Page 2 - 3508 Controller

Notes 10 MC19-1.20  For a Step segment Holdback delays continuation to the next segment until the step target is reached. The Holdback Type can be s

Page 3 - Contents

Notes 11 MC19-1.20 Program number On P10 or P25 models press up  or down  to select a new program number. The display will show that new pro

Page 4

Notes 12 MC19-1.20 5.6 Running a Program The current program can be started from the home display by pressing RUN/HOLD. Or by pressing page  once, t

Page 5 - 2 Basic Operation

Notes 13 MC19-1.20 Program setpoint (PSP) Current Segment Type. Step and Call segments are instant, so are only flashed on the screen, unl

Page 7

Notes 15 MC19-1.20 5.13 Program Example 1 The following sequence of entries creates and runs the program shown graphically below. 1. Turn th

Page 8 - 4 Setting up the controller

Notes 16 MC19-1.20 5.14 Program Example 2 The following sequence of entries creates and runs the program shown graphically below. 1. Turn the Setpoi

Page 9 - 5 Programming

Basic Operation 17 MC19-1.20 6 Options Because options can be ordered in a variety of combinations and for a variety of purposes, exact instru

Page 10 - MC19-1.20

Notes 18 MC19-1.20 To change the address value access the level 2 list. In level 2 press the page key until the COMMS parameter is displayed. Press

Page 11 - 11

Basic Operation 19 MC19-1.20 7 Navigation Diagrams 7.1 Operator Level 1 - No Program Running Current program page Programming Page Home di

Page 12

3508 Controller 2 MC19-1.20 Note: For the purposes of this manual the Eurotherm 3504 range is functionally identical to the 3508 range.

Page 13 - 13

Notes 20 MC19-1.20 7.2 Operator Level 1 - Program Running Current program pageProgramming PageHome displayCustomer IdentityProgStat program nrSee

Page 14

Notes 21 MC19-1.20 7.3 Supervisor Level 2 Current program pageProgramming PageHome displayCustomer ID nrControl Output H

Page 15

Basic Operation 22 MC19-1.20 8 Controller Fault 8.1 Fault Code Diagnostic Table Error Code Explanation Actions S.br Temperature sensor failure

Page 16 - P25 only

Notes 23 MC19-1.20

Page 17 - 6 Options

Notes Copyright © 2014 Carbolite Limited Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)143

Page 18

3508 Controller 3 MC19-1.20 Contents 1.1 Using This Manual 4 1.2 PID Control 4 2 Basic Operation 5 2.1 Controller Layout (Home display) 5 2.2

Page 19 - 7 Navigation Diagrams

3508 Controller 4 MC19-1.20 Introduction to the Controller and Manual 1.1 Using This Manual This manual aims to explain how to set up and operate the

Page 20

Basic Operation 5 MC19-1.20 2 Basic Operation 2.1 Controller Layout (Home display) 2.2 Keys A/Man - Disabled RUN/HOLD - Runs,

Page 21 - 21

Basic Operation 6 MC19-1.20 3 Quick Start Guide 3.1 Operation as a simple controller When switched on the controller goes through a short test

Page 22 - 9 Glossary of Terms

Controller Fault and Glossary of Terms 7 MC19-1.20 2. Press up  to select level 2. After a short pause the display will show “Access Pass code”. 3.

Page 23 - 23

Basic Operation 8 MC19-1.20 4 Setting up the controller Before using the controller (or during its lifetime) certain parameters may have to be s

Page 24

Basic Operation 9 MC19-1.20 5 Programming 5.1 Programming Notes Programs can be created in Level 1 or Level 2 of the 3508. For the P10 and P25,