3.0 OPERATION
The instructions for operating the temperature controller are given in a separate manual.
If the furnace is fitted with a time switch, see also the supplementary manual MS03.
3.1 Operating Cycle
If the furnace has an auxiliary heater, see also section 3.2.
The furnace is fitted with an Instrument switch. The switch cuts off power to the control circuit.
There is also a Heater switch which can be used to disconnect power to the elements.
Connect the furnace to the electrical supply.
Operate the instrument switch to activate the temperature controller. The controller becomes
illuminated and goes through a short test cycle. Adjust the temperature controller (see controller
manual).
Overtemperature option only. If the overtemperature controller has not yet been set as required, set
it and activate it according to the instructions in the appropriate manual.
Switch on the Heater switch, located on the instrument panel.
Unless a time switch is fitted and is off, the furnace starts to heat up. The Heat light(s) glow
steadily at first and then flash as the furnace approaches the desired temperature or a program
setpoint.
Unless a time switch is fitted and is off, the furnace starts to heat up. The Heat light(s) glow
steadily at first and then flash as the furnace approaches the desired temperature or a program
setpoint.
Overtemperature option only. If the overtemperature trip operates then an indicator in the
overtemperature controller flashes, and the heating elements are isolated. Find and correct the
cause before resetting the overtemperature controller according the instructions supplied.
To switch off power to the heating elements, use the Heater Switch .
To turn the furnace off, set the Instrument switch to it’s off position; the controller display will go
blank. If the furnace is to be left off unattended, isolate it the electrical supply.
3.2 Auxiliary heater (if fitted)
The auxiliary heater is controlled by an independent temperature controller, which is affected by
the same instrument switch as the main controller. Note that a time switch (if fitted) may not
switch off the auxiliary heater, depending on other options ordered.
Set the auxiliary heater's temperature controller after setting the main controller, as given in
section.
Auxiliary heater controller navigation diagram:
Home
List
Access
List
20.0
ACCS
OP
codE
*C
*C
output power,
read only
measured temperature;
use
/
to access
setpoint
for factory access to
lists and parameters
not available to the
operator