Carbolite CDF 15/1b User Manual Page 15

  • Download
  • Add to my manuals
  • Print
  • Page
    / 16
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 14
CDF
MF56 1.03 15
8.0 FUSES & POWER SETTINGS
8.1 Fuses
F1-F3: Refer to the circuit diagrams.
F1
Internal
supply fuses
Fitted if supply cable fitted.
Fitted on-board to some types of EMC filter.
16 Amps:
38mm x 10mm type F
F2
Auxiliary
circuit fuses
Fitted on-board to some types of EMC filter.
May be omitted up to 10 Amp/phase supply rating.
2 Amps glass type F
On board: 32mm x 6mm
Customer
fuses
Recommended if cable fitted.
See rating label for
amperage; see table below
for fuse rating.
Model
phases
Volts
Supply Fuse Rating
CDF 15/1b
1-phase
200-230
13A/ph
Page view 14
1 2 ... 10 11 12 13 14 15 16

Comments to this Manuals

No comments